Journal of Cardiovascular Medicine and Therapeutics

All submissions of the EM system will be redirected to Online Manuscript Submission System. Authors are requested to submit articles directly to Online Manuscript Submission System of respective journal.
Reach Us +44-1518081136

Short Communication - Journal of Cardiovascular Medicine and Therapeutics (2020) Volume 4, Issue 3

Expandable Polyurethane Stent Valve®, implanted by catheter. Results of physical, hydrodynamic and experimental tests

A Expandable Polyurethane Stent Valve®, (EPSV), with a special design for pediatric patients, may be an option for biological prostheses, whose calcification or mismatch, in child’s development, shorten their durability.
Methods: An expandable chrome - cobalt stent, was applied polyurethane, for the formation of three cuspids, without sewing, it was sumitted to:1- In vitro tests: I- Physical test. Universal tests of samples of PU crimped and non-crimped was performed: A- Strength versus deformation (stretching). B- Scanner for surface for mechanical properties. II- Hydrodynamic test. Using a pulsatile flow, to register: valvular area, pressure gradient and valve regurgitation. 2- In vivo test: III- Experimental: Ten sheep were submitted to implantation of expandable polyurethane stent valve by catheter, in pulmonary position, using right ventricle approach. Expansion diameter: Group A: 22mm (7 cases) and Group B: 18mm (3 cases).
Results: I - In vitro tests: PU showed to be a thermoplastic structure with high deformation, resistant to crimping and elongation. The hydrodynamic test showed low gradients and absent or trivial regurgitation of the prosthesis. 2- In vivo tests: After 5th months of follow-up, a 3D echocardiographic study, was performed in eight survival sheep and showed: satisfactory hemodynamic performance, with no significant transvalvular gradient (M = 6.60 mmHg), absent or trivial valvular
regurgitation, absent leak and free of calcification. Five survival implanted sheep are well after 24 months of follow up.
Conclusions: Monitoring of favorable results, confirms that the expandable PU stent valve, can be implanted in pulmonar position in growing children, below 7 years, during conventional surgery and above 7 years, by peripheral vessel.
Author(s): Miguel Maluf

Abstract PDF

Get the App