Journal of Biochemistry and Biotechnology

All submissions of the EM system will be redirected to Online Manuscript Submission System. Authors are requested to submit articles directly to Online Manuscript Submission System of respective journal.

Research Article - Journal of Biochemistry and Biotechnology (2017) Volume 0, Issue 0

Determination of acarbose in tablets by attenuated total reflectance Fourier transform infrared spectroscopy

Attenuated total reflectance Fourier transform infrared (ATR- FTIR) spectroscopy method was developed for the quantitative determination of acarbose in tablets. Two calibration curves were generated. The first calibration curve was based on peak height at 2928.54 cm-1 with correlation coefficient of 0.9987. The second calibration curve was based on peak area which range from 2953.39-2851.04 cm-1 with correlation coefficient of 0.9978. The FTIR method was validated according to International Conference on Harmonization (ICH) guidelines. The method was precise over a range of 40%-100% w/w with % RSD values less than 2 for intra-day and interday precision. The accuracy of the method was expressed as the percentage recovery of the spiked acarbose standards with a mean recovery of 106.9%. The limit of detection and limit of quantification values were 0.004% and 0.012 % w/w, respectively. The validated method was used for the quantification of acarbose in tablets with good results.

Author(s): Ali Saoud, Gabriel Akyirem Akowuah*, Omotayo Fatokun, Ahmad Mariam and Shaik Ibrahim Khalivulla

Abstract Full Text PDF

Get the App